The calibration of the tear tensile tester usually includes the following steps: 1. Prepare a standard dynamometer: First of all, you need to prepare a standard dynamometer, such as a standard weight with known accuracy or a dynamometer specially used for calibration of tear-tensile force testers.2. Install the dynamometer: Install the standard dynamometer on the sensor of the tear force tester to ensure that the installation is stable and does not loosen. 3. Adjust the zero point: Adjust the zero point of the tear-pull force tester when the dynamometer is not loaded with any load to ensure that the display value is zero. 4. Load standard loads: Use standard dynamometers to load known standard loads, such as 10 Newtons, 20 Newtons, etc. 5. Read the display value: After loading the standard load, read the display value of the tear tensile tester and record the difference between the actual display value and the standard load. 6. Calculate the error: Calculate the error of the tearing force tester according to the difference between the standard load of the load and the actual value. If the error is within the acceptable range, it means that the calibration is successful. If the error is too large, the tester needs to be further checked or adjusted. Need to noteIt means that the calibration of the tear-tensile tester should be operated by professionals to ensure the accuracy and safety of the calibration. At the same time, in order to ensure the accuracy of the test results, it is also necessary to maintain and maintain the tear-tensile force tester regularly.